CALIBRATION OF UV/ VISIBLE SPECTROPHOTMETER

Ensure that the connection of the instrument are proper.
Control of absorbance
Potassium dichromate solution:
Dry a quantity of potassium dichromate by heating to constant weight at 130°C. Weigh & transfer accurately
a quantity not less than 57.0 mg & not more than 63.0 mg to 1000ml volumetric flask. Dissolve & dilute in sufficient 0.005M H2SO4 to produce 1000ml.
Measure the absorbance of potassium dichromate solution at the wavelength at the wavelengths given below.
Calculate value of A (1% 1cm) for each wavelength.
A(1% 1cm) = Absorbance X 10000 / Weight of potassium dichromate in me.
Acceptance criteria
Wavelength (nm)
A [1% 1cm]
Limit
235.0
124.5
122.9 – 126.2
257.0
144.0
142.8 -145.7
313.0
48.6
47.0 – 50.3
350.0
106.6
105.6 – 108.2

Stringent limit are adopted from IP & BP for acceptance criteria
Resolution power
Record the spectrum of a 0.02% v/v solution of toluene in hexane in the range of 260nm to 420 nm (before use check the hexane for transmittance, using water as a blank between 260nm to 420nm & use only if transmittance is not less than 97%).
Acceptance criteria:
The ratio of the absorbance at the maximum at about 269 nm to that at the minimum at about 266nm is not less than1.5.
Limit of stray light
Prepare a 1.2 % w/v solution of potassium chloride in water.
Measure absorbance of the above solution at 198.0, 199.0, 200.0, 201.0, 202.0 nm using water as blank.
Acceptance criteria:
Absorbance is greater than 2.
Wavelength accuracy, Resolution & Baseline flatness (inbuilt test)
Attach printer directly to the instrument (instead of computer).
Go to > MODE press F3 key i.e. Maintenance.
Press  ‘1 ‘.
Press Start/ Stop Key.
After screen changes, ensure that nothing is kept in the optical path & press the Start / Stop key again.
Prints will come after all the three tests are over.
Acceptance criteria:
For wave length accuracy : at 656.1 ± 0.3 nm & at 486.0 ± 0.3 nm.
Resolution: 1.0 nm or less.
Baseline flatness: ±0.002Abs.
Frequency: Once in a month.
Maintenance / Repairs
When the instrument does not comply with the requirement / tolerance range specified above, the instrument should be labeled “OUT OF CALIBRATION” and should be repaired / serviced.
After repair / maintenance, calibration the instrument.
Reference:
Indian Pharmacopoeia
British Pharmacopoeia
Instruction manual of UV-1700.





CALIBRATION REPORT OF UV/ VISIBLE SPECTROPHOTOMETER
SOP Reference -                                                                                               Date-
Instrument No-                                                                                                                Frequency – Once in a month
Make -                                                                                                                  Model –
  1. Control of Absorbance:
Potassium dichromate solution:
Gross Wt:                                                                                            Balance :
Tare Wt :
Net Wt :                                                               ________ ml with 0.005M sulphuric acid.
Dissolve and dilute 57.0 to 63.0 mg of potassium dichromate in sufficient 0.005M sulphuric acid to produce 1000.0 ml.
Wavelength
Absorbance
A (1%, 1cm)
Tolerance
235.0


122.9 to 126.2
257.0


142.8 to 145.7
313.0


47.0 to 50.3
350.0


105.6 to 108.2
Calculation:
1%                          Absorbance X 10000
A1cm     =      --------------------------------------------------------   =  -------------------  =
         Weight of potassium dichromate in mg
(Limit : Not less than 1.5)
Remark : Satisfactory /Not satisfactory.
Performed By:                                                                                  Checked By:
Date:                                                                                                     Date:

  1. Resolution Power:
0.02% v/v toluene in hexane:
………………… ml   -> 100.0 ml with hexane / …………… ml  -> 100.0 ml with hexane
Absorbance at 269 nm = ……………………..
Absorbance at 266 nm = ……………………..
                Absorbance at 269 nm
Ratio = ----------------------------------- = ------------------- =
                Absorbance at 266 nm
(Limit : Not less than 1.5)
Remark: satisfactory / Not satisfactory.
Performed BY:                                                                  Checked By:
Date:                                                                                     Date:
  1. Limit of stray light:
Potassium chloride solution
Gross Wt :_________________                                                                              Balance : __________
Tare Wt : __________________            
Net Wt : ___________________                           _________ml with water.
Dissolve and dilute accurately 1.2g potassium chloride in sufficient water to produce 100.0 ml.

Sr. No
Wavelength (nm)
Absorbance
Tolerance
1
198.0

Absorbance is greater than 2
2
199.0

3
200.0

4
201.0

5
202.0

Remarks : Satisfactory / Not satisfactory.
Performed By:                                                                                  Checked By:
Date:                                                                                                     Date:
  1. Wavelength accuracy (In built test)
Sr No
Wavelength Standard (nm)
Wavelength Observed (nm)
Difference
Limit (nm)
1
656.1


±0.3
2
486.0


±0.3
Remark: Satisfactory / Not satisfactory.
Performed By:                                                                                  Checked By:
Date:                                                                                                     Date:
  1. Resolution (In built test)
Sr No
Limit
Observation
1
1.0 nm or less
Nm
Remark: Satisfactory / Not satisfactory.
Performed By:                                                                                  Checked By:
Date:                                                                                                     Date:
  1. Baseline flatness (In built test)
Sr.No
Limit
Observation
1
± 0.002 Absorbance

Remark : Satisfactory / Not satisfactory.
Performed By:                                                                                  Checked By:
Date:                                                                                                     Date:

Conclusion: Instrument working is Satisfactory / Not Satisfactory.
Next Calibration Due On:

Performed By:
Date:
Checked By:
Date: