1.
Principle
1.1
This
appendix presents some information on the characteristics that should be
considered during validation of analytical methods. Approaches other than those
specified in this appendix may be followed and may be acceptable. Manufacturers
should choose the validation protocol and
procedures most suitable for testing of their product.
procedures most suitable for testing of their product.
1.2
The
manufacturer should demonstrate(Through validation) that the analytical
procedure is suitable for its intended purpose.
1.3
Analytical
methods, whether or not they indicate stability, should be validated.
1.4
The
analytical method should be validated by research and development before being
transferred to the quality control unit when appropriate.
2.
General
2.1
There
should be specifications for both, materials and products. The tests to be
performed should be described in the documentation on standard test methods.
2.2
Specifications
and standard test methods in pharmacopoeias (“pharmacopoeial methods”), or
suitably developed specifications or test methods (“non-pharmacopoeial
methods”) as approved by the national drug regulatory authority may be used.
2.3
Well-characterized
reference materials, with documented purity, should be used in the validation
study.
2.4
The
most common analytical procedures include identification tests, assay of drug
substances and pharmaceutical products, quantitative tests for content of
impurities and limit tests for impurities. Other analytical procedures include
dissolution testing and determination of particle size.
2.5
The
results of analytical procedures should be reliable, accurate and reproducible.
The characteristics that should be considered during validation of analytical
method are discussed in paragraph 6.
2.6
Verification
or revalidation should be performed when relevant, for example, when there are
changes in the process for synthesis of the drug substance; changes in the
composition of the finished product; changes in the analytical procedure; when
analytical methods are transferred from one laboratory to another; or when
major pieces of equipment instrument change.
2.7
The
verification or degree of revalidation depend on the nature of the change(s).
2.8
There
should be evidence that the analysts, who are responsible for certain tests,
are appropriately qualified to perform those analyses (“analyst proficiency”).
3.
Pharmacopoeial
methods
3.1
When
pharmcopoeial mentods are used, evidence should be available to prove that such
methods are suitable for routine use in the laboratory (verification).
3.2
Pharmacopoeial
method used for determination of content or impurities in pharmaceutical
products should also have been demonstrated to be specific with respect to the
substance under consideration (no placebo interference).
4.
Non-pharmacopoeial methods
4.1
Non-pharmacopoeial
methods should be appropriately validated.
5.
Method validation
5.1
validation
should be performed in accordance with the validation protocol. The protocol
should include procedures and acceptance criteria for all characteristics. The
results should be documented in the validation report.
5.2
Justification
should be provided when non-pharmacopoeial methods are used if pharamcopoeial
methods are available. Justification should include data such as comparisons
with the pharamcopoeial or other methods are available.
5.3
Standard
test methods should be described in detail and should provide sufficient
information to allow properly trained analysts to perform the analysis in a
reliable manner. As a minimum, the description should include the
chromatographic conditions (in the case of chromatographic tests), reagents
needed, reference standards, the formulae for the calculation of results and
system suitability tests.
6.
Characteristics of analytical procedures
6.1
Characteristics
that should be considered during validation of analytical methods include:
-
Specificity
-
Linearity
-
Range
-
Accuracy
-
Precision
-
Detection
limit
-
Quantitation
limit
-
Robustness
6.1.1
Accuracy
is the degree of agreement of test results with the true value, or the
closeness of the results obtained by the procedure to true value. It is normal
established on samples of the material to be examined that have been prepared
to quantitative accuracy. Accuracy should be established across the specified
range of the analytical procedure.
6.1.2
Precision
is the degree of agreement among individual results. The complete procedure should
be applied repeatedly to separate, identical samples drawn from the same homogeneous batch of material. It should be measured by the scatter of
individual results from the mean (good grouping) and expressed as the relative
standard deviation (RSD).
6.1.2.1
Repeatability
should be assessed using a minimum of nine determinations covering the
specified range for the procedure e.g. three concentrations/ three replicates
each, or a minimum of six determinations at 100% of the test concentration.
6.1.2.2
Intermediate
precision expresses within-laboratory variations (usually on different days,
different analysts and different equipment).
6.1.2.3
Reproducibility
is assessed, a measure of intermediate precision is not required.
6.1.3
Robustness
(or ruggedness) is the ability of the procedure to provide analytical results
of acceptable accuracy and precision under a variety of conditions. The results
from separate sample are influenced by changes in the operational or
environmental conditions. The results from separate samples are influenced by
changes in the operational or environmental conditions. Robustness should be
considered during the development phase, and should show the reliability of an
analysis when deliberate variations are made in method parameters.
6.1.3.1
Factors
that can have an effect on robustness when performing chromatographic analysis
include;
-
Stability
of test and standard sample and solutions;
-
Reagents
(e.g. different suppliers);
-
Different
columns (e.g. different lots and/or suppliers);
-
Extraction
time;
-
Variations
of pH of a mobile phase;
-
Variations
in mobile phase composition;
-
Temperature;
and
-
Flow
rate.
6.1.4
Linearity
indicates the ability to produce results that are directly proportional to the concentration of the analyte in samples. A series of samples should be prepared
in which the analyte concentration span the claimed range of the procedure. If
there is a linear relationship, test results should be evaluated by appropriate
statistical methods. A minimum of five concentrations should be used.
6.1.5
Range
is an expression of the lowest and highest levels of analyte that have been demonstrated to be determinable for the product. The specified range is
normally derived from linearity studies.
6.1.6
Specificity
(selectivity) is the ability to measure unequivocally the desired analyte in
the presence of components such as excipients and impurities that may also be
expected to be present. An investigation of specificity should be conducted
during the validation of identification tests, the determination of impurities
and assay.
6.1.7
Detection
limit (limit of detection) is the smallest quantity of an analyte that can be detected, and not necessarily determined, in a quantitative fashion. Approaches
may include instrumental or non-instrumental procedures and could include those
based on:
-
Visual
evaluation;
-
Signal
to noise ratio;
-
Standard
deviation of the response and the slope;
-
Standard
deviation of the blank; and
-
Calibration
curve.
6.1.8
Quantitation
limit (Limit of Quantitation) is the lowest concentration of an analyte in
sample that may be determined with acceptable accuracy and precision.
Approaches may include instrumental or non-instrumental procedures and could
include those based on:
-
Visual
evaluation;
-
Signal
to noise ratio;
-
Standard
deviation of the response and the slope;
-
Standard
deviation of the blank; and
-
Calibration
curve.
6.2
Characteristics
(including tests) that should be considered when using different types of
analytical procedures are summarized in Table 1.
Table 1
Characteristics to consider during analytical validation
|
||||
Type of Analytical procedure
|
Identification
|
Testing for impurities
|
Testing for impurities
|
Assay
- dissolution (measuremen only)
- Content/potency
|
Characteristics
|
Quantitative test
|
Limit tests
|
||
Accuracy
|
-
|
+
|
-
|
+
|
Precision
Repeatability
Intermediate precision
|
-
-
|
+
+
|
-
-
|
+
+
|
Specificity
|
+
|
+
|
+
|
+
|
Detection limit
|
-
|
-b
|
+
|
-
|
Quantitation limit
|
-
|
+
|
-
|
-
|
Linearity
|
-
|
+
|
-
|
+
|
Range
|
-
|
+
|
-
|
+
|
- Characteristic is
normally not evaluated;
+ Characteristic
should normally be evaluated.
-a In
cases where a reproducibility study has been performed, intermediate
precision is not needed.
-b May be
needed in some cases.
|
6.3
System
suitability testing:
System suitability testing is an
integral part of many analytical procedures. The tests are based on the concept
that the equipment, electronics, analytical operations and samples to be
analysed constitute an integral system that can be evaluated as such. System
suitability test parameters that need to be established for a particular
procedure depend on the type of procedure being evaluated, for instance, a
resolution test for an HPLC procedure.
No comments:
Post a Comment