CALIBRATION
OF TOP PAN BALANCE
Calibration procedure:
Internal Calibration
It shall be performed as per the manufacturer’s
instructions.
Drift Check
Note down the 10 measurements of the 10 mg weight in
the performance check log. (Attachment No. 01)
Variation in the observed weight from the mean value
does not exceeds ± 0.2 mg.
The 10mg weight shall meet the performance check
critical of the mass value i.e. 0.1% of actual mass value.
E.g. For all the 10 measurements of 10 mg weight,
variation in the weighings cannot exceed 0.01 mg.
Performance Check
After the Auto Calibration Put 1 mg, 2 mg, 5 mg, 10 mg
and 20 mg weights individually. The measurement shall within the 0.1% of actual
mass value of the individual weight as given in the performance check log.
(Attachment 02)
Measurement uncertainty check
The measurement of uncertainty shall be carried out by
using 10 mg weight.
Put the external weight of 10 mg on the pan and note
the 10 measurements. (Attachment no.03)
Calculate the measurement of uncertainty as follows.
Measurement
uncertainty = Standard Deviation X 3
Actual mass value
Measurement of uncertainty shall be Not More Than
0.001.
Calibration frequency: It shall
be performed daily, after any maintenance, after the relocation and after the power
failure of the balance.
Each manufacturer supplies the maximum loading
condition of the balance. The calibration of the balance shall be designed in
such a way that the performance check weights cover the entire loading range of
the balance. Lower higher load limits shall be checked for the performance
check. Before Calibration make sure that the level bubble is in the center of
the indicator and if internal calibration facility is available in the balance
then it shall also be performed daily before performance check of the balance.
Always use a check weight appropriate for the
particular balance. E.g. a 100 – mg check weight might be selected for a
balance has a load limit of 150 mg. Thus the operator must know the maximum
capacity of the balance to select the correct check weight.
Attachment-1: Performance Check Log of Microbalance
Instrument:________________ Code No:________________
Make :________________ Model
:________________
Date
|
Level Indicator check
|
1 mg
|
2 mg
|
5 mg
|
10 mg
|
20 mg
|
Weight box no. and valid upto
|
Done by
|
Checked by
|
Remark
|
Std.Wt.Cal No. Þ
|
||||||||||
Limit Þ
|
Bubble should be in Center
|
±0.1% of Std. Weight ( to
)
|
±0.1% of Std. Weight ( to
)
|
±0.1% of Std. Weight ( to
)
|
±0.1% of Std. Weight ( to
)
|
±0.1% of Std. Weight ( to
)
|
||||
Attachment – 2: Auto Performance Check log of
Microbalance
Instrument:________________ Code No:________________
Make :________________ Model
:________________
Check level Indicator (Bubble should be in center) :
__________________________________
(Paste
Performance Check Chart Here)
Calibrated By :_________________ Date :_________________
Checked By
:_________________ Date
:_________________
Attachment – 3: Measurement Uncertainty Check of
Microbalance
Instrument:________________ Code No:________________
Make :________________ Model :________________
Date
|
Std.Wt
|
Std. Wt. Cal.No.
|
Observation
|
Observation of Uncertainty Less than or equal to
0.001
|
Weight box no. and valid upto
|
Done by
|
Checked by
|
Remark
|
|||||
10 mg
|
|||||||||||||
Calculation
Measurement
uncertainty = Standard Deviation X 3
Actual Mass Value
No comments:
Post a Comment