CALIBRATION OF BROOKFIELD VISCOMETER

Procedure:
Ensure that all the connection of the instrument is proper.
Turn the viscometer ON.
Take out the viscosity standard solution in 500ml beaker, adjust the temperature to 25° C. Attach the
required spindle, adjust required speed & measure the viscosity, not e down the constant reading.
Viscosity standards, speeds, spindles required & tolerance limits are as follows:
Sr. No
Viscosity of standard in cps
Actual viscosity of standard in cps
Spindle No.
Entry Code
RPM
Tolerance (in cps)
Minimum
Maximum
1.
5.00
4.70
ULA
00
60
4.55
4.85
2.
50.00
48.40
LV1
61
60
46.9
4909
3.
5000.00
4820.00
LV3
63
10
4651.83
4988.17

Acceptance criteria
Viscosity should be within the tolerance limits.
Do not transfer the viscosity solution to the original container back after use.
Precautions
Take due care to avoid any contamination of the standard solutions.
Avoid air bubble entrapment during transfer of standard solutions.
Frequency
Once in 6 months.
Maintenance / Repairs:
When the instrument do not comply with the requirement/ tolerance range specified above, the instrument should be labeled “OUT OF CALIBRATION” and should be repaired / served.

 

 

QUALITY ASSURANCE DEPARTMENT

CALIBRATION REPORT OF BROOKFIELD VISCOMETER

Reference-                                                                                         Date
Frequency – Once in 6 months                                                  Instrument No-
Make – Brookfield Engineering Laboratories, Inc.              Model – LVDV – II + Pro
Take out the viscosity standard solution in 500ml beaker, adjust the temperature to 25°C & attach the required spindle, adjust required speed & measure the viscosity, note down the constant reading.
Sr. No
Viscosity of standard in cps
Actual viscosity of standard in cps
Observed Viscosity CPS
Tolerance (in cps)
Remarks
(S / NS)
Minimum
Maximum
1.
5.00
4.70

4.55
4.85

2.
50.00
48.40

46.9
4909

3.
5000.00
4820.00

4651.83
4988.17

Conclusion: Calibration is satisfactory (S) / not satisfactory (NS)
Next Calibration Due On:
Done By:
Date:
Checked By:
Date: