1. PURPOSE
The purpose of this SOP
is
1. To defined the operating procedure.
3. Cleaning procedure and maintenance.
2. SCOPE
This SOP is applicable
for operation, handling cleaning and maintenance of the high performance liquid
chromatography. Make: Jasco Model: UV-975, PU-980
3. RESPONSIBILITY
To organize the job
training program for implementation of SOP.
To monitor overall
activities in adherence to the compliance of approved SOP.
To arrange necessary
servicing and maintenance.
To check the recorded
documents for its correctness, accuracy and completeness.
Section in charge
To operate the
instrument as per SOP.
To record the
observations as per SOP.
To maintain the
documents as per SOP.
4. SAFETY WARES
Wear apron, cap and
company foot ware.
5. START UP PROCEDURE
5.1 Clean the high pressure liquid
chromatography (HPLC) system and surroundings.
5.2 Ensure that the system consists of
following equipment
5.3 Select the prescribed column and
connect in the specified direction i.e. in the direction of flow arrow / sign
marked on the column, column outlet end is connected to the inlet of the
detector, which in turn is connected to ‘Drain’ through tubing.
5.4 Ensure that the mobile phase flow is
in following direction.
5.5 Ensure that UPS is working.
5.6 Connect the power supply and switch
ON the mains.
5.7 Instrument is ready for calibration
or use.
6. OPERATIONAL PROCEDURE
6.1 PUMP
6.1.1
Ensure
that, the instrument is calibrated and start up procedure is followed.
6.1.2
Turn
the power switch, located at the lower left of the front panel, to ‘ON’
position.
6.1.3
The
LED display glows on and it follows diagnostic tests.
6.1.4
Wait
till initialization is completed.
6.1.5
Press
“ EDIT / ENTER” or ‘2’ key. The flow value on display starts blinking then
press EDIT / ENTER key.
6.1.6
Set
the flow at 0.2ml /min, using the numerical key on the panel, and again press
“EDIT / ENTER” key.
6.1.7
Press
‘5’ key. The value of P.MAX will start blinking. Set the value at 500; press
“EDIT / ENTER” key.
6.1.8
Press
‘6’ key. The value of P.MIN will start blinking. Set the value at 0; press
“EDIT / ENTER “ Key.
6.1.9
Ensure
that end of inlet tubing of pump is in the reservoir containing at least 50 ml
filtered and degassed methanol (HPLC Grade).
6.1.10 Open the purge valve by rotating the
knob in anticlockwise direction.
6.1.11 Press ‘PUMP’ key. The orange pilot
lamp indicates that the pump is on.
6.1.12 The methanol flows through the pump
and is drained out from the drain tube connected to the outlet of purge valve.
6.1.13 Allow to run methanol for 5 minutes
or till any air bubble produced in the tube is drained out.
6.1.14 Press ‘PUMP’ key. The pilot lamp is
off.
6.1.15 Close the purge valve by rotating it
in clockwise direction.
6.1.16 Press ‘2’ key. And set the flow at
0.2 ml/min. Press “EDIT /ENTER” key. Press ‘PUMP’ key.
6.1.17 The valve on the monitor screen
indicates the actual pressure.
6.1.18 Gradually increase the flow with the
interval of 0.2 ml / min to 1.0 ml / min.
6.1.19 Allow to run for 10 minutes.
6.1.20 Gradually decrease the flow with the
interval of 0.2 ml/ min; to 0.0 ml / min.
6.1.21 Press ‘ PUMP’ key. Remove the tubing
from methanol and place it in the reservoir containing filtered and degassed
(sonicated) water for HPLC.
6.1.22 Follow the instructions 11 to 18.
6.1.23 Allow to run for at least 20 minutes.
Then decreases the flow. And press ‘PUMP’ key.
6.1.24 Replace the water with reservoir
containing mobile phase.
6.1.25 Follow the instructions 11 to 18 and
increase the flow to the specified flow rate.
6.1.26 Allow the run at the specified flow
rate for at least 15 minutes or as specified in the individual monograph so as
equilibrate the system with the mobile phase.
6.1.27 The system is now ready for
injection.
6.1.28 Inject the samples as per individual
specification.
6.1.29 After completing all the injections,
switch off the detector, make entry in usage log.
6.1.30 Gradually decrease the flow with the
interval of 0.2 ml/min to 0.0 ml/min.
6.1.31 Press ‘PUMP’ key. Remove the tubing
from mobile phase and place it in the reservoir containing filtered and
(degassed) sonicated water for HPLC.
6.1.32 Follow the instructions 6.1.11 to
6.1.18.
6.1.33 Allow to run for at least 240
minutes. Then decrease the flow and press ‘PUMP’ key.
6.1.34 Replace the water with reservoir
containing methanol.
6.1.35 Follow the instructions 6.1.11 to
6.1.18.
6.1.36 Allow to run for at least 20
minutes. Then decrease the flow and press ‘PUMP’ key.
6.1.37 Switch off the pump.
6.1.38 Switch off the mains.
6.2 DETECTOR
6.2.1
Turn
the power switch, located at the lower left of the front panel, to ‘ON’
position.
6.2.2
The
LED display glows on and it follows diagnostic tests.
6.2.3
Wait
till initialization is completed.
6.2.4
Press
“EDIT /ENTER” or ‘1’ key. The value of the wavelength on display starts
blinking.
6.2.5
Enter
the specified wavelength using the numerical keys.
6.2.6
Press
“EDIT /ENTER” key for 2 times or press ‘2’ key. The value of the range on display
starts blinking.
6.2.7
Enter
the specified range using or keys. Again press EDIT / ENTER key.
6.2.8
The
absorbance value will be displayed on the screen.
6.2.9
The
system is now ready for analysis.
6.3 INTEGRATOR (NT BASED SOFTWARE)
6.3.1
Switch
on the computer in which the software is installed.
6.3.2
Wait
till the initialization is complete.
6.3.3
Activate
the software undergoes start up procedures.
6.3.4
Wait
till the software undergoes start up procedures.
6.3.5
Enter
the specified ‘OPERATOR’ Name, ‘USER’ name and then click ‘OK’.
6.3.6
The
screen now displays the software window with the selected ‘USER’>
6.3.7
Click
on ‘RUN’ key, then monitor Base Line, select the specified system.
6.3.8
Enter
the appropriate information in the displayed chart. i.e. Run Name, No., Run
Length and run information.
6.3.9
Click
‘RUN’ key, the screen will display, chromatogram showing ‘Waiting’ status. The
system is now ready for analysis.
7. ANALYSIS PROCEDURE
7.1 Use only HPLC Grade solvents for
analysis.
7.2 Use only calibrated, clean,
dry glass apparatus.
7.3 All the solutions which are entering
the HPLC System i.e. water, mobile phase, etc. must be filtered through 0.4 µ
membrane filter paper, and properly degassed.
7.4 In case of mobile phase containing
buffer, it is essential to give water washing to the column prior and after
running the mobile phase.
7.5 Prepare the sample and standard
solutions as mentioned in the specifications.
7.6 Ensure that the HPLC system is
equilibrated, and ready to start.
7.7 Take a clean, dry syringe.
Rinse it with water in case of mobile phase containing buffers, or with the
mobile phase.
7.8 Remove red cap of Rheodyne injector
port.
7.9 Ensure that the injector is on
‘Load’ position.
7.10 Suck water / mobile phase in the
syringe, ensure that no air bubble is trapped in the syringe.
7.11 Inject water / mobile phase in the loop. Rinse
the port with water for 5 times.
7.12 Inject mobile phase or the
solvent system used for sample preparation and turn the injector downward i.e.
in ‘Inject’ position, the channel on the screen gets activated and shows
‘Run in Progress’ signal.
7.13 Turn the injector port on ‘Load’
position.
7.14 After
completing the injections again rinse the loop with water and then with
Methanol.
7.15 Preserve the column in
Methanol.
No comments:
Post a Comment